Remote Control Commands
R&S
®
SMA100B
467User Manual 1178.3834.02 ─ 05
Parameters:
<RMode> SINGle | CONTinuous
*RST: CONTinuous
Example:
SENS:SWE:FREQ:RMOD SING
Selects single measurement
Manual operation: See " Execution - Power Analysis " on page 234
:SENSe[:POWer]:SWEep:FREQuency:SPACing[:MODE] <Mode>
Selects the spacing for the frequency power analysis.
Parameters:
<Mode> LINear | LOGarithmic
*RST: LINear
Example:
SENS:SWE:FREQ:SPAC:MODE LIN
Sets linear spacing of the sweep
Manual operation: See " Spacing - Power Analysis " on page 233
:SENSe[:POWer]:SWEep:FREQuency:STARt <Start>
Sets the start frequency for the frequency mode.
Parameters:
<Start> float
Range: 0 to 1E12
*RST: 1E6
Example:
SENS:SWE:FREQ:STAR 2.0GHZ
Sets a sweep start at 2 GHz.
Manual operation: See "Min - Power Analysis" on page 231
:SENSe[:POWer]:SWEep:FREQuency:STEPs <Steps>
Sets the number of measurement steps for the frequency mode.
Parameters:
<Steps> integer
Range: 10 to 1000
*RST: 200
Example:
SENS:SWE:FREQ:STEP 500
Sets 500 steps
Manual operation: See " Steps - Power Analysis " on page 232
Power Sensor Measurement Subsystems