Remote Control Commands
R&S
®
SMA100B
479User Manual 1178.3834.02 ─ 05
Usage: Event
Manual operation: See "Save To Ref" on page 228
:SENSe[:POWer]:SWEep:POWer:REFerence:DATA:POINts?
Queries the number of points from the reference curve in "Power" measurement.
Return values:
<Points> integer
Range: 10 to 1000
Example:
SENS:POW:SWE:POW:REF:DATA:POIN?
Queries the number of points from the reference curve in power
mode.
Usage: Query only
:SENSe[:POWer]:SWEep:POWer:REFerence:DATA:XVALues <XValues>
Sets or queries the x values of the two reference points, i.e. "Power X (Point A)" and
"Power X (Point B)" in "Power" measurement.
Parameters:
<XValues> string
Example:
SENSe:POW:SWE:POW:REF:DATA:XVAL -15DBM,20DBM
Sets the x value of reference "Point A" to -15 dBm, and the
value of "Point B" to 20 dBm.
Manual operation: See "Pow (X) / Pow (Y)" on page 228
:SENSe[:POWer]:SWEep:POWer:REFerence:DATA:YVALues <YValues>
Sets or queries the y values of the two reference points, i.e. "Power Y (Point A)" and
"Power Y (Point B)" in "Power" measurement.
Parameters:
<YValues> string
Example:
SENSe:POW:SWE:TIME:REF:DATA:YVAL -30,10
Sets the y value of reference "Point A" to -30 dBm, and the
value of "Point B" to 10 dBm.
Manual operation: See "Pow (X) / Pow (Y)" on page 228
:SENSe[:POWer]:SWEep:POWer:RMODe <RMode>
Selects single or continuous mode for measurement mode power in power analysis.
Parameters:
<RMode> SINGle | CONTinuous
*RST: CONTinuous
Power Sensor Measurement Subsystems