Remote Control Commands
R&S
®
SMA100B
633User Manual 1178.3834.02 ─ 05
Manual operation: See " Current Level" on page 172
[:SOURce<hw>]:POWer:MODE <Mode>
Selects the operating mode of the instrument to set the output level.
Parameters:
<Mode> CW | FIXed | SWEep
CW|FIXed
Operates at a constant level.
CW and FIXed are synonyms.
To set the output level value, use the command [:
SOURce<hw>]:POWer[:LEVel][:IMMediate][:
AMPLitude].
SWEep
Sets sweep mode.
Set the range and current level with the commands:
[:SOURce<hw>]:POWer:STARt and [:SOURce<hw>]:
POWer:STOP,
[:SOURce<hw>]:POWer:MANual.
*RST:
CW
Example: See Example "Setup an RF frequency or power sweep"
on page 643
Manual operation: See "State (RF level sweep)" on page 171
[:SOURce<hw>]:POWer:POWer <Power>
Sets the level at the RF output connector.
This value does not consider a specified offset.
The command [:SOURce<hw>]:POWer[:LEVel][:IMMediate][:AMPLitude]
sets the level of the "Level" display, that means the level containing offset.
See "RF frequency and level display with a downstream instrument" on page 63.
Parameters:
<Power> float
Level at the RF output, without level offset
Range: See data sheet
Increment: 0.01
Default unit: dBm
Example:
SOURce1:POWer:POWer 15
Sets the level at RF output
Manual operation: See "Amplitude" on page 68
SOURce Subsystem