Page | 43 DRM-3000 Operating Manual
Calibration of the external detectors is achieved using a designated USB to Detector cable. The
calibration factor is saved in the detector and therefore the calibration of the external detectors
can be achieved independently of the DRM-3000 Meter.
The calibration procedure for the internal and external wide range detectors is identical to the
calibration procedure of the TelePole.
The RMVC Software automatically creates a network connection and therefore needs to run in the
Administrator mode. After the installation procedure, Set the Compatibility to Administrator as
shown below: