Both cubitainers are positioned inside the RPI in designated drawers.
The upper position is indicated as the "Top" concentrate cubitainer (1).
The lower position is indicated as the "Bottom" concentrate cubitainer (2).
The RPI-Check bottle can be found inside the door at the right side of the instrument (3).
During normal operation it is possible to produce reagent until both cubitainers are empty.
After placing a new cubitainer, the data of this specific cubitainer is stored in the instrument: The ID of the
cubitainer and the batch information of the reagent is logged for quality control purposes. This data is also
visible in the software .
Always use concentrated reagent cubitainers that has been acclimatized for at least 24 hours to
environment temperature.