If more errors are active, it will be indicated:
10.1 - No water
-Possible stop of water supply, check water connection/supply.
-WPU-error, check display of the WPU-unit, in case of no water, check tubing and water supply.
10.2 - Concentrated reagent information is not recognized
If after placing a new concentrated reagent cubitainer the data is not recognized (error F55), the instrument
will not start production. This data has to be manually added in the
cubitainers menu.
Troubleshooting
MRN-184-EN Version 6 (05-01-2022)
69