12
Error list
Description Possible cause(s) How to solve?
F01 Mix vessel will not fill (process error) Issue with WPU
Instrument failure
LT: Check WPU, drain mix vessel,
restart process, if not solved:
Fatal error, call for service
F02 Mix vessel will not fill (hardware error) Issue with WPU
Instrument failure
LT: Check WPU, drain mix vessel,
restart process, if not solved:
Call for service
F03 Mix vessel will not empty Instrument failure Call for service
F05 Internal test ECT unit timeout Issue with the ECT module Call for service
F06 Air bubbles detected in control process Bubbles being pumped RPI-
Check bottle is (almost) empty
RPI-Check tubing/pickup is
broken
Replace RPI-Check bottle
Check pick up tube
If not solved:
Call for service
F07 Temperature is out of range Temperature too high or low Wait for temperature change
F08 Cell constant out of range ECT unit issue Perform Control cycle with another
RPI-Check
Check RPI-Check tubing
Download and send control data to
FSE for further analysis
F09 Fluid left in mix vessel (above low float) DI valve leaking
lower float suddenly broken
Call for service
F10 Temperature lab water out of range Temperature of DI out of range Check AFS inlet temperature.
If not solved, call for service
F11 ECT module is not stable ECT Sensor error Restart
Control cycle
If not solved, call for service
Error list
MRN-184-EN Version 6 (05-01-2022)
89