11.1 - Maintenance items & Maintenance schedule RPI
Item Order nr. Interval Performed by
Concentrated Reagent customer
specific
If empty Operator
RPI-Check A0020487 (12
pcs)
If empty Operator
Quarterly maintenance RPI (with parts from Quarterly Maintenance Kit A0020395):
Replacement endpoint filters &
tubing (4 pcs)
QLV0400013 3 months Lab technician
Replacement RPI-Check tubing
(1 pcs)
A0020834 3 months Lab technician
Replacement concentrate
tubing (2 pcs)
A0022758
(ECO)
3 months Lab technician
Replace Concentrate pickup
tubes (2 pcs)
A0022751
(ECO)
3 months Lab technician
Vent filter SQWLV040002 3 months Lab technician
Maintenance of WPU unit:
Replacement WPU filters see AFS manual 5/6 months Lab technician
Chlorine treatment of WPU see AFS manual upon instrument request~3 months Lab technician
pH treatment of WPU see AFS manual upon instrument request~1 year Lab technician
Yearly maintenance:
Scheduled RPI maintenance
with use of Yearly Maintenance
Kit
with parts from
A0020396
1 year FSE
Scheduled WPU maintenance
(AFS-24)
see AFS manual 1 year Merck FSE
Maintenance
MRN-184-EN Version 6 (05-01-2022)
73