Power Supply
The power connection has to be accessible at any time, the power cord is functioning as disconnecting
device.
Should the instrument emit unusual smell or smoke, turn the main switch OFF immediately and unplug
the power cable. Using the instrument further carries the risk of fire, electric shock or personal injury.
Contact your Service Representative. Do not disconnect the power for any other reason.
Do not touch the electric circuits inside the instrument particularly with wet hands, as there is a risk of
electric shock.
This instrument must be connected to a power outlet of the correct voltage. Please note that the
instrument must be earthed. Use only the original delivered main power cord or exact equivalent with
wire thickness of 3x1.0 mm2 (AWG 18) according local regulations.
Avoid damage to the power cable. Do not place any appliances on the power cable or pull on the power
cable.
9.1 - Avoidance of infection
The internal fluid containers may never be opened by the user, but only by trained service personnel. During
prolonged use of the RPI, harmful bacteria (f.e. legionella) may grow inside the containers. Before the
containers are opened, the disinfection procedure must be performed (see section Maintenance-Cleaning for
details).
Operational precautions, limitations and hazards
MRN-184-EN Version 6 (05-01-2022)