11.5 - Replacing Pickup tubes/End point filters
Only to be performed by a trained Lab Technician or FSE. Be aware of liquid spilling, use a cloth
to prevent liquid dripping in the instrument.
Replacing Pickup tubes:
1. Enter software and log in with the applicable credentials.
2. Go to Maintenance and select "RPI".
3. Press Replace Pickup tubes/End filter, the following message is given.
4. Press YES to confirm and start actual replacement.
5. Press YES after the message "Instrument will be offline."
5. Replace Pick-up tubes of both concentrated reagent cubitainers.
Maintenance
82
MRN-184-EN Version 6 (05-01-2022)