1. Ensure connected analyzers are placed in an operational mode where reagent consumption will not
occur.
2.
If not done yet: R
emove the top panel by loosening the three M4 bolts(with Torx T10) at the back,
lifting up and fully removing from the instrument.
3. Detach the end filter tubing, take care of leaking fluids and remove the old filters.
4. Place new A0017999 End point filter assemblies (filters+tubing), connect tubes according the numbers.
5. Place and fasten the top panel and place side panels (if still open) if no other maintenance action is
needed.
6. Connected analyzers can resume consumption.
Maintenance