F26 CAN address error Wiring incorrect or electronics
failure
Call for service
F27 ECT measurement precision too low ECT error Call for service
F28 ECT module doesn't have valid control
data
Fluid bottle is empty
ECT module issue
(re)start Control cycle
Place RPI-Check bottle
Call for service
F29 No conductivity detected after starting
control cycle
Pickup tube/Peristaltic pump
tube issue
Replace the peristaltic tubing
Restart the control cycle
F50 Temperature of 1x reagent in the
storage vessel out of range
External temperature too high
or too low
Automatic Drain storage and wait for
restart production
F51 NFC tag for RPI-Check not successfully
read and door is closed
No RPI-Check bottle installed
NFC tag damaged
Control orientation incorrect
Place new bottle or move control
bottle
F52 RPI-Check content too low for next
cycle
After control cycle, too little
RPI-Check for next control cycle
Place new control bottle
F53 Automated cycle not completed due to
insufficient content
Not sufficient RPI-Check when
starting control cycle
Place new control bottle and start
control cycle
F54 One concentrate cubitainer empty Placement of new cubitainer is
advised
Place new cubitainer
F55 One concentrate cubitainer not detected No cubitainer installed
NFC tag damaged
Cubitainer orientation incorrect
Reposition cubitainer
If still not read:
Add NFC tag manually
F56 Both concentrate cubitainers empty/not
detected
Production is on hold Place two new cubitainers
F57 Open life one concentrate cubitainer
expired
Open life of the concentrate
RPI-Check is expired
Check cubitainers and bottle, replace
applicable
F58 Open life both concentrate cubitainers
expired
Open life of the concentrate
RPI-Check is expired
Check cubitainers and bottle, replace
applicable
Error list
MRN-184-EN Version 6 (05-01-2022)