Press “ MANL ZERO ” to enter below menu:
Press numeric keys to enter the new zero and press “ SURE ZERO ” to validate it.
3.2.2.2
3.2.2.2
3.2.2.2
3.2.2.2 Span
Span
Span
Span Calibration
Calibration
Calibration
Calibration
Span calibration is the adjustment between the difference in Belt Scale indication and a
known weight standard . Span is also called as weigh span, i.e. indication response to change
of standard weight on the load receptor. The purpose of span calibration is to remedy the
difference between the reading and the standard weight on the load receptor.
There are two different system calibration modes: simulated calibration and material
calibration.
Three simulated span calibration methods are available: R-cal, test weights and test chains.
Simulated calibration needs caculating “ Simulated constant ” , see section 3.3.10 for its
calculation in details. When performing span calibration, the calibration method should be
firstly selected, and the relevant resistor or test weights / chain weights parameters should be
entered into the integrator.
The following Span Calibration menu is accessible by selecting “ SPAN CALIBRATION ”
from the Calibration Menu, then press “ ENT ” :
2010-11-09 12:00:00 Scale#1
8 000000 t
Old zero : [ 888 ]
New zero :
Main: Manual zero mode !
Input zero
:
BGIN EXIT ENTR ENTR
ZERO ZERO ZERO
2006-3-21 12:00:00 Scale#1
CALIBRATION
2.1 ZERO CALIBRATION
2.2 SPAN CALIBRATION
2.3 MATL CALIBRATION
Press ↑↓ key, then press ENT!
MAIN
MENU