3.10
Revised 3/04/11
SOFTWARE CONFIGURATION
5.
Selectable print head resolution settings are 360 or
180 dpi. With slide surfaces not suitable for 360 dpi
resolution, printing results will be poor. For such slides,
180 dpi should be selected.
6.
The
Paper/Output
menu item is for the unload station
only.
Same Tray: Slides are deposited into a tray until
0
the tray is full.
Job in new tray: Each print job starts with a new
0
tray.
New Tray: Only for special applications-do not
0
select this option under standard Windows
programs.
NOTE:
The instrument will ignore the parameters selected
in the paper/output menu if the slide bin is being used.
7.
The
Print Quality
menu is used to select whether a
print is to be applied onto a slide once (normal) or
twice (overstrike). Faint prints can be improved using
the overstrike option. The Upside down feature turns
rotates the prints 180°. Overstrike can also be chosen
for the upside down option.
NOTE:
It is recommended that the overstrike feature be
used only when necessary, as the ink consumption greatly
increases.