Revised 3/30/2010 B.1
Appendix B
WARNINGS AND CAUTIONS
Warnings
Do not bring fire sources close to the
system
. This system uses organic solvent that may
ignite upon contact with fire. Do not bring a fire source
within 2 m of the system.
Do not bring fire sources close to the
activated carbon filters.
The activated carbon
filters may explode upon contact with fire. Do not bring a
fire source within 2 m of the activated carbon filters.
Do not disassemble or modify the system.
The system may malfunction or cause an accident.
Do not open covers and doors unless
necessary. If the operator must open a
cover and insert a hand into the system,
do so only after confirming that the
system has stopped and the internal
components can be operated safely
.
If a door must be opened and a basket set,
do so only after confirming that the LOAD
lamp is on and the internal components
can be operated safely.
The system is designed
to stop when a cover is opened. However, the actual
operation may not stop until the series of steps comprising
the operation is completed. Touching the system while it is
still operating may result in injury.
When opening/closing any cover, be
careful not to pinch your hand.
Be careful not
to pinch your hand between the fixed and moving parts of
a cover or between a cover and a door, etc. Your hand
may be pinched and injured.
Do not wet the system. Wetting any part of the
system other than those specified in Section 5, “Care of
the Instrument” or other applicable sections may cause
fire or electric shock due to leak current.
Do not operate the switches and
connect/disconnect the power plug with
wet hands. Doing so may result in electric shock.
Connect/disconnect the power plug by
holding the plug.
Connecting/disconnecting the
power plug by holding any other part other than the plugs
may result in fire or electric shock.
Do not forcibly bend, pull, twist or loop the
power cable.
Use of a damaged power cable may
result in fire or electric shock. Should any damage or other
abnormality on the power cable or plug be found, contact
the Sakura dealer.
Handle reagents (organic solvents) with
due care.
Some reagents (organic solvents) used by
this system may be toxic or harmful to the human body.
Accordingly, handle these reagents with due care to
prevent them from spilling or scattering. Dispose of
reagent properly by following the applicable regulation or
guideline in your country or region.