52
Setting procedure of Control panel.
When use MCO-5GC(the automatic switcher of CO
2
gas supply line),it is necessary to set the
MCO-5M automatic gas switching mode according to the following procedures.
Description of operation Key
operated
Indication after operation
1 Turn the power switch ON.
-
The current chamber temperature is
displayed.
2 Press CAL key for 5 seconds.
CAL
The left digit in the temperature indicator
is flashed.
3 By pressing key and key,
set the figure to F03.
4 Press ENT key.
ENT
In CO
2
density indicator,
"000" is displayed and the
right digit is flashed.
5 By pressing key and key,
set the figure to 384
6 Press ENT key.
ENT
The current chamber temperature is
displayed.
7 Press CAL key for 5 seconds.
CAL
The left digit in the temperature indicator
is flashed.
8 By pressing key and key,
set the figure to F08
9 Press ENT key.
ENT
In CO
2
density indicator,
"0**" is displayed and the
right digit is flashed.
10 By pressing key and key,
set only the left digit figure to 1.
NOTE:Don't change center
digit figure and right
digit figure.
"*" means "0" or "1".
11 Press ENT key.
ENT
Setting of MCO-5GC is finished.
The current chamber temperature is
displayed.
CO
2
gas supply line indicator A is lighted.
F 0 3
F 0 3 3 8 4
F 0 8
F 0 3 0 0 0
F 0 8 0 * *
F 0 8 1 * *