EasyManua.ls Logo

Sanyo MCO-5M - Page 56

Sanyo MCO-5M
57 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
53
Procedure for automatic gas switching operation check.
When the installation of MCO-5GC is completed, check the automatic gas switching operation
according to the following to the following procedures.
procedure
1 Stop the gas supply of cylinder A and use cylinder B only for CO
2
gas supply.
2 Turn on the power switch of MCO-5M and set 37decC and 0%.
Turn on CO
2
gas supply line indicator with CO
2
gas supply line switching key.
3 Wait for approx.1 minute until CO
2
control is available.
(It takes about 1 minute until CO
2
control is effective after power ON.)
4 Set the CO
2
density to 5% and check CO
2
inject lamp lights.
NOTE: CO
2
inject lamp may not light depending on the temperature, humidity or
CO
2
density in the chamber. In this case, set the CO
2
density 1% or more
higher than the displayed value.
5 Check the E01(CO
2
gas cylinder empty alarm) and buzzer turn on about
2-7minutes later from the CO
2
inject lamp's turning on.
6 Check the CO
2
gas density indicator reaches to the set value 5% and keep it
stably.
NOTE: CO
2
density control is interrupted when automatic calibration of CO
2
sensor
is activated and the decimal point of CO
2
density indicator blinks.
7 This is the end of check procedure.
Press the buzzer key to disappear “E01” and select the cylinder A with CO
2
gas
supply select key.
Set the CO
2
density to 0% and open the door to release the CO
2
gas completely.
After the chamber temperature and humidity are stable, set the CO
2
density to
desired value.
Refer to the MCO-5M instruction manual for usage of MCO-5GC
7FB6P104047000

Other manuals for Sanyo MCO-5M

Related product manuals