17
7. Function mode
F00 Automatically revert to PV display
F01 SV
H
(high temp. alarm SV) setting
F02 SV
L
(low temp. alarm SV) setting
F03 Battery accumulating time display
F04 Automatically revert to PV display
F05 Compressor delay time setting
F06 Service code input (code: 384)
F07 * Temperature Zero Adjustment
F08 * Automatically revert to PV display
F09 * Factory test mode …… Unavailable
F10 * Factory test mode …… Unavailable
F11 * Factory test mode …… Unavailable
F12 * Display of temperature of temp. sensor
F13 * Automatically revert to PV display
F14 * Display of temperature of Comp. sensor
F15 * Automatically revert to PV display
F16 * Automatically revert to PV display
F17 * Model code setting (Initialization of non-volatile ROM and memory)
F18 * Automatically revert to PV display
F19 * Automatically revert to PV display
F20 * Automatically revert to PV display
F21 Communication ID set
F22 Communication mode set
F23 * Automatically revert to PV display
F24 Remote alarm terminal output
F25 Ring-back time set
F26 * “000” is displayed…Unavailable
F27 * “000” is displayed…Unavailable
F28 * Display of delay time of permission for measuring running rate (2 hrs timer)
F29 * Display of delay time of permission for measuring running rate (8 hrs timer)
F30 * ROM version is displayed
F31 * Automatically revert to PV display
F32 * Notice for condensing fan motor accumulating time
F33 * Automatically revert to PV display
F34~F39 Unused
Setting
procedure:
In PV display, keep pressing key over 5seconds to display “F00”.
Input the desired function code with key and key.
Press SET key to be function mode available.
* … You should input service code (384) in F06 beforehand.
To cancel service code, input “000” in F06 or turn the power off.
F00: <Purpose> Simply passing through if entered by mistake.
<Operation> Press SET key in “F00” displayed to revert to PV display.
F01: <Purpose> SV
H
(high temp. alarm SV) setting
<Operation> Input F01 and press SET key to display “010” (initial value).
Set the value in the range “005”~“020” with key. Press SET key
to store the value and revert to PV display.
F02: <Purpose> SV
L
(low temp. alarm SV) setting
<Operation> Input F02 and press SET key to display “-10” (initial value).
Set the value in the range “-05”~“-20” with key. Press SET key to
store the value and revert to PV display.
F03: <Purpose> Battery accumulating time is displayed
<Operation> Input F03 and press SET key to display alternately F03 with “00.0”
(in case battery used for a month or less). Press SET key to revert to
PV display.