Information on Weighing Electro statically Charged Samples 
 
lf you are using a balance with a 0.1 mg readability and electro stati-
cally charged glass or plastic Containers with a diameter larger than 
the weighing pan, you should utilize our antistatic pan  
(Ø 130 mm - see "Accessories") instead of the Standard weighing 
pan. 
 
 
 
 
 
 
 
 
Leveling the Balance Using the Level Indicator (depends on type 
of model) 
 
At the point of use, level the balance using the leveling feet 
(8)
 so that 
the air bubble is centered within the circle of the level indicator 
(10)
. 
 
To raise the balance, lower the leveling feet (turn clockwise). 
To lower the balance, raise the feet (turn counter-clockwise).