Automatic printing (PROTOC menu item 7.15):
When the menu item (7.15.2) is active, the first weight value that exceeds the
minimum load is printed.
If the menu item is also activated for automatic taring, it is only tared when the
minimum load is exceeded. In this case, an automatic printout would only be
generated when the second weight value exceeds the minimum load.
Main scale: first platform displayed on start-up (Combics 2 only)
You can select the weighing platform to be displayed first when Combics is turned
on in the Setup menu under “UTILIT." (menu item 8.11.).
Entering a tare weight using a barcode scanner (Combics 2 only)
The tare weight of the container can be entered via a barcode scanner. To do this,
the Tare setting must be activated in the menu under Setup/Barcode.
The value is applied and saved automatically, the t key does not have to be
pressed. The content of the tare memory can be displayed in Info mode (I key).
Entering the wRef application parameter using a barcode scanner
(Combics 2 only)
wRef application parameters can be entered via a barcode scanner. To do this, the
wref setting must be activated in the menu under Setup/Barcode. The value is
applied and saved automatically, the r key does not have to be pressed.
Entering Identifier using a barcode scanner (Combics 2 only)
Identifiers can be entered via a barcode scanner.
ID1: To do this, the ID1 setting must be activated in the menu under Setup/
Barcode. The value is applied and saved automatically, the d key does not
have to be pressed.
ID2 to ID6: To do this, the HEADER setting must be activated in the menu under
Setup/Barcode. Then press the d key until the desired ID entry
appears, scan the barcode and save.
Displaying the content of the identifiers:
– d key
Scanning barcodes directly
You can directly scan a barcode using the barcode scanner.
Menu setting: Setup / BARCODE/ INPUT
The barcode can contain the following codes:
- I for write identifier
- T for save tare memory
- R for write reference weight
- A for activate product data memory
Examples: “I4Anton" = write the character string to ID 4: Anton
“TC1" = write 1 Kg to the preset tare memory.
“C" = unit: Kilograms
“B" = grams
“D" = carat,
etc.
“RC0.0023" = write 0.0023 kg as the reference weight
“A1" = load product data memory 1
Menu setting: Setup /BARCODE / HEADER
The characters read from the barcode are shown in the weight value display.
Operating Instructions Combics Indicators 41
Operation