Adjustment/Configuration counter for standard scales
Purpose Automatic recording of changes to adjustment and weighing parameters using two
independent counters. The values remain saved for the life of the component.
t To display both counters, press and hold the ( key for longer than 2 seconds.
y The “Configuration counter" is then shown in the weight display for 3 seconds
(identified by a P). Then the “Adjustment counter" is displayed for another
3 seconds (identified by a C). After 6 seconds, the information display turns off
automatically.
Adjustment counter features: – Counter limited to 9999
– Counter at “C 0000" for hardware commissioning
– Counter cannot be reset
– Counter is updated automatically when:
– linearization, calibration/adjustment is successful
– user calibration, adjustment or linearization weight is changed (menu 1.18.)
– When the following parameters are changed:
Function of the q key (menu item 1.9.)
Zero setting range (menu item 1.11)
Tare/zero at power on (menu item 1.12)
The above parameters are reset to factory settings (menu item 9.1.1)
Configuration counter features: – Counter limited to 9999
– Counter at “P 0000" for hardware commissioning
– Counter cannot be reset
– Counter is updated automatically when:
– When the following parameters are changed:
Installation location (menu item 1.1.)
Application filter (menu item 1.2.)
Stability range (menu item 1.3.)
Taring (menu item 1.5)
Auto zero (menu item 1.6.)
Weight unit 1 (menu item 1.7.)
Weight unit 2 (menu item 3.1.)
Weight unit 3 (menu item 3.3.)
The above parameters are reset to factory settings (menu item 9.1.1)
– Switching the k key to or from a 10-fold higher resolution
– Turning the application automatic taring on/off
(menu item 3.7.)
– The application parameters are reset to factory settings
(menu item 9.1.1)
42 Operating Instructions Combics Indicators