EasyManuals Logo

Sartorius CP2202S-0CE User Manual

Sartorius CP2202S-0CE
50 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #22 background imageLoading...
Page #22 background image
22 |
Adjusting External Linearity
Checking the Linearity
Note: The linearity must be checked according to the Standard Operating Procedure
WKD-038.
- Check the linearity of the balance/scale for the entire weighing range in 4-g
steps.
- Compare the values displayed with the tolerance ranges given in the „Table of
Calibration/Adjustment Data“ on pages 42-43.
- If the errors exceed the permissible tolerance ranges, the linearity must be
adjusted.
Adjusting the Linearity
Caution! You can adjust the linearity on CP balances/scales only using the SARTORIUS MC1
Server Software (Version 4.9 and later) or with the SARTOCAS program.
- Activate the BPI mode (see page 15).
- Adjust the linearity with the SARTORIUS MC1 Server (version 4.9 and later) or
with the SARTOCAS program for PCs and laptops.
(Please see the program description for instructions on this automatic proce-
dure!).
Adjusting Linearity

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius CP2202S-0CE and is the answer not in the manual?

Sartorius CP2202S-0CE Specifications

General IconGeneral
BrandSartorius
ModelCP2202S-0CE
CategoryScales
LanguageEnglish

Related product manuals