EVO1S1 | SPO1S1 Operating Instructions 15
Installation
4 Installation
The scale must be disconnected from the power supply for all assembly work.
4.1 Unpacking and Equipment Supplied
t Open the packaging, making sure to remove all parts carefully.
t After unpacking the device, check it immediately for any external damage.
t If you detect any damage, proceed as directed in Chapter “7 Care and
Maintenance,” page 29.
t Save the box and all parts of the packaging for any future transport. All cables
should be unplugged when transporting.
The following parts are included in the equipment supplied:
Model name PMA.Evolution PMA.HD
Large weighing pan: d 233 mm
x –
Small weighing pan: d 180 mm
– x
USB connection cable x x
AC adapter with country-specific mains adapter Optional Optional
Installation instructions x x
4.2 Selecting a Setup Location
Select the right setup location:
− Set up the device on a stable, even surface that is not exposed to vibrations.
− Maintain free access to the device at all times.
Choose a location that is not subject to the following negative influences:
− Heat (heater or direct sunlight)
− Drafts from open windows, AC systems and doors
− Extreme vibrations during weighing
− Heavy “traffic areas” (personnel)
− Extremely high humidity
− Electromagnetic fields
− Extremely dry air
Acclimatization
Condensation from humidity can form on the surfaces of a cold device when it
is brought into a warm area. You should therefore let a device acclimatize for
approximately two hours disconnected from its power source before reconnecting
it to the supply voltage.
Seal on Scales Verified for Use in Legal Metrology:
Legislation requires that a seal be affixed to verified scales. On Sartorius devices, this
seal takes the form of a sticker with the “Sartorius” logo. If the seal is removed, the
validity of verification will become void and you must have your scale re-verified.
The verification supplied here is for verified weighing instruments for use in the EEA.
Please keep it in a safe place.