20 EVO1S1 | SPO1S1 Operating Instructions
Operation
6 Operation
6.1 Switching the Device On/Off
Switching On
t Briefly press the j key (on/standby).
y An automatic self-test runs. This ends when the display shows 0.0 g.
t If another value is displayed: Set the scale to zero via the 0 key.
Switching Off
t Press and hold the j key for several seconds (on/standby).
y The scale switches to standby mode.
6.2 Locking the Keypad/Weight Display
Locking
t Briefly press the j key (on/standby) to lock the keypad and turn off the weight
display.
y The padlock symbol H flashes for several seconds.
t Press the padlock symbol H while flashing to activate the lock.
y The keypad/weight display is locked and the padlock symbol lights up continually.
Unlocking
t Press the padlock symbol H to deactivate the lock.
t Enter the password (if set) (see Chapter 6.7.2.6, page 27).
y The lock is deactivated.
6.3 Zeroing/Taring
t Briefly press the 0 key.
6.4 Calibration/Adjustment
The scale needs to be re-leveled and then adjusted each time its setup location is
changed.
Warm-up Time
To ensure accurate results are delivered, the scale must warm up for at least
30 minutes after initial connection to the power supply.
Only then will the device have reached the required operating temperature.
t Wait approx. 30 minutes after connecting the scale to the power supply before
adjusting the scale and weighing.
Using a Scale Verified for Use in Legal Metrology:
Ensure that there is a warm-up time of at least 24 hours after initial connection
to the power supply.
0
30

Table of Contents

Other manuals for Sartorius EVO1S1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius EVO1S1 and is the answer not in the manual?

Sartorius EVO1S1 Specifications

General IconGeneral
BrandSartorius
ModelEVO1S1
CategoryScales
LanguageEnglish

Related product manuals