22 EVO1S1 | SPO1S1 Operating Instructions
Operation
6.5.2 Weighing with Two Decimal Places (Not for Verified Models)
A menu setting is required for weighing with two decimal places (see
Chapter “6.7.2.4 Activating the Toggle Key (Not for Verified Models),” page 26).
t Place the empty paint can on the weighing pan.
t Briefly press the 0 key to zero.
y The display shows “0.0 g.”
t Press the toggle key Z.
y The display shows “0.00 g.”
t Add the first component “205.50 g.”
t Read off the weight when the stability symbol (in this example) “g” is displayed.
t Add the other components until the desired weight (formula) is reached.
t Remove the filled paint can from the weighing pan.
When the scale is tared and the second decimal place with a resolution of 0.05 g is
activated via the toggle key Z, weighing up to 999.95 g with two decimal places
can be performed.
Values over 999.95 g can only be weighed using one decimal place.
Danger of Scale Damage!
Never close a paint can using a hammer while it is still on the weighing pan.
t When closing the paint can, place it on a firm, stable surface.
6.6 Applications
Note for Scales for Use in Legal Metrology:
All application programs can be selected on scales verified for use in legal metrology.
Calculated values can be marked using the following characters:
Percent = %
Computed values = o, l symbol
0.0
0.00
205.50
213.50
593.85
1414.2

Table of Contents

Other manuals for Sartorius EVO1S1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius EVO1S1 and is the answer not in the manual?

Sartorius EVO1S1 Specifications

General IconGeneral
BrandSartorius
ModelEVO1S1
CategoryScales
LanguageEnglish

Related product manuals