Setting the Drying Parameters
Purpose
Adapt the moisture analyzer to the special requirements of products.
Parameters can be configured individually for each program.
Drying Parameters (Overview)
ο Factory setting
√ User-defined setting
Program ο P1
memory P2
(PROGRAM) P3 Heating program ο Standard drying Temperature 105°C 40...230°C
MA30 mode Temperature 105°C 40...160°C
1
)
Standby ο Off
temperature
2
) On Temperature 40°C 40...100°C
Initial weight ο Off
On Target 5.0 g 0.2 g...35.0 g
Start analysis With stability
when Enter pressed
ο Without stability
when cover is closed
End of analysis ο Fully automatic
Semi-automatic Loss/24 s 5 mg 1...20 mg
Timer mode Time 15.0 min 0.1...99.9 minutes
Display mode ο Moisture (%L)
Dry weight (%R)
Ratio (%LR)
Residual weight (g)
Residual weight (g/kg)
Print ο Off
intermediate results On Interval 0.1 min 0.1...10.0 minutes
Factory settings No
Yes
1
) Marked in Info/Status line by an asterisk (“*”); e.g. “*105
o
C”
2
) Not when using the “MA30 mode” heating program
24
Operating the Moisture Analyzer
Factory setting
Factory setting
Factory setting