23
Applying a Sample to the Sample Pan
§ Apply the sample to the sample pan in a thin, even layer
(height: 2 to 5 mm, weight: 5 to 15 g); otherwise:
a sample applied unevenly will result in a non-uniform distribution of heat
a sample will not be dried completely
the analysis time will be unnecessarily extended
the sample burns or a crust/ skin forms on its surface as a result of a very thick layer
the crust makes it difficult or impossible for moisture to escape from the sample during
the drying process
uncertain and unknown quantity of moisture remains in the sample
§ Apply liquid samples, pasty samples or samples that can melt to a glass fiber filter
(order no. 6906940); you will obtain the following advantages:
uniform distribution due to the capillary effect
liquids prevented from beading together and forming drops
with larger surfaces, the moisture can evaporate faster
considerably more convenient than the “sea-sand method”
When drying samples containing sugar, a crust or skin can form and seal the surface.
A glass fiber filter is especially helpful in such cases. The moisture can evaporate down-
wards through the surface of the filter. You can prevent or limit the crust/ skin formation
if you place the glass fiber filter on top of the sample.
§ Cover solid, heat-sensitive samples with a glass fiber filter (order no. 6906940); you will
obtain the following advantages:
gentle heating, because of the sample surface is shielded from excessive heat
higher temperature setting can be selected
uniformity of the sample surface
fast evaporation of the moisture
excellent reproducibility for samples containing fat
Avoiding the Formation of Crust/Skin
You can add “solvents” to the sample to prevent the formation of crust/skin during the
analysis runs. The weight of a solvent you apply to the sample will not influence the final
result of an analysis.
§ Once you have closed the sample chamber, open it again within 2 seconds after you hear
a beep tone.
§ Apply a solvent to the sample
§ Close the sample chamber and start the analysis run as usual