Error codes are displayed in the main display or text line for 2 seconds.
The program then returns automatically to the previous status.
Display Cause Solution
No segments appear on the display No AC power is available Check the AC power supply
The power cord is not plugged in Plug in the power cord
Fuses are defective Replace the fuses
H The load exceeds the weighing capacity Unload the pan support
L or ERR 54 The pan support is not in place Place the pan support
on the weighing system
ERR 01 Data output not compatible with Change the configuration in
> Display range output format the Setup menu
ERR 02 Calibration/adjustment condition not met, e.g., Calibrate only when zero is displayed
Cal. n. possible not tared Press the TARE function key
the pan support is loaded Unload the moisture analyzer
ERR 03 Calibration/adjustment could not Allow the moisture analyzer to warm up
Cal/adj. interrupt be completed within a certain time again and repeat the adjustment process
ERR 30 Interface port for printer output Have the port setting changed
Print fct. blocked is blocked by Sartorius Customer Service
ERR 31 External device not ready to send Transmit XON, then CTS
Print fct. blocked (interface handshake interrupted; XOFF, CTS)
ERR 53 No temperature compensation Contact the Sartorius Service Center
No meas. value
ERR 101 Key is stuck Release key
Key pressed when switching on or contact your local
the moisture analyzer Sartorius Service Center
“Checkerboard” pattern displayed u key was pressed when turning
continuously on the moisture analyzer, or is stuck
ERR 320 Operating program memory is wrong Contact your local
Sartorius Service Center
ERR 340 Operating parameter (EEPROM) Contact your local
is wrong Sartorius Service Center
ERR 342 Operating parameter (EEPROM) is Contact your local
wrong except for adjustment parameters Sartorius Service Center
NO WP Weighing system is defective Contact your local
Sartorius Service Center
blocked Function blocked None
xxxxx too low Input wrong (with any Follow the instructions
xxxxx too high application program), e.g., for the application programs
alphabetic input not allowed
Troubleshooting: see next page
44
Error Codes

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius MA150C-000115V1 and is the answer not in the manual?

Sartorius MA150C-000115V1 Specifications

General IconGeneral
BrandSartorius
ModelMA150C-000115V1
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals