Step Key (or Instruction) Display/Printout
The weighing system is prepared
for the calibration procedure
When the system is ready, the display shows
10. Place the weight on the weighing system Apply 30-g calibration
(in this example, 30.000 g) weight
Minus sign: weight too low
Plus sign: weight too high
No sign: weight ok
At the end of calibration,
the display shows
11. If no adjustment is required, y
select END in the function line u
and confirm --------------------
13.08.2001 16:15 *)
Model MA45-000230V
*) Date and time shown only when Ser. no. 99992581
using the YDP02 or YDP03 printer Ver. no. 01-43-01
from Sartorius ID
--------------------
External calibration
W-ID
Nom. + 30.000 g
Diff. - 0.002 g
--------------------
13.08.2001 16:16 *)
Name:
--------------------
12. If necessary, adjust the u
weighing system
At the end of adjustment,
the display shows
and a report is printed (see next page)
33
30000
PROGRAMM
WGH.SYS CAL./ADJ.Verzög. 2sec
D
g
TARASETUP
30000
PROGRAMM
WGH.SYS CAL./ADJ.Verzög. 2sec
D
g
TARASETUP
0002
END
WGH.SYS CAL./ADJ.Verzög. 2sec
D
g
ADJUSTSETUP
0002
ENDE
WGH.SYS CAL./ADJ.Verzög. 2sec
D
g
JUST.SETUP
30000
CAL.
WGH.SYS CAL./ADJ.Verzög. 2sec
S
g
TAREEND