Specifications
Model MA45
Dryer functions:
Heating element Ceramic IR heater
Temperature range 40 – 230°C
Temperature increments adjustable in 1°C increments
Temperature adjustment with YTM03MA temperature adjustment set
Weighing Functions:
Weighing capacity 45 g
Readability 1 mg, 0.01% moisture content
Repeatability, average (%) sample weight 1–5 g: 0.2%
sample weight 5 g: 0.05%
External calibration weight 30 g (± 0.3 mg)
(with an accuracy of at least...)
Sample pan dimensions 90 mm
Drying parameters:
Drying programs Standard; MA30 mode
Drying time 0.1 to 99.9 minutes
Number of programs 3
Shutoff criteria Fully automatic, semi-autom. (1 – 20 mg/24 s),
timer mode (1 ƒ 99.9 Min.)
Display of analysis results Moisture, dry weight, RATIO, residual weight (g or g/kg)
Minimum sample weight 0.100 g
Analyzer (hardware):
Dimensions (W ƒ D ƒ H) 213 ƒ 320 ƒ 181 mm
Net weight, approx. 5.1 kg
Voltage 230 V or 115 V (depending on model variant), –15% ... +10%
Frequency 48 – 60 Hz
Fuses 2 (neutral conductor/phase), 6.3 AT, 5 ƒ 20 mm (internal)
Operating temperature range +10 ... +30°C
Power consumption 700 VA max.
Built-in interface RS-232C
Format: 7 bit ASCII, 1 start bit, 1 or 2 stop bits
Parity: Odd or even
Transmission rates: 150 to 19,200 baud
Handshake: Software or hardware
48
Overview

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius MA150C-000115V1 and is the answer not in the manual?

Sartorius MA150C-000115V1 Specifications

General IconGeneral
BrandSartorius
ModelMA150C-000115V1
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals