EasyManua.ls Logo

Sartorius MA50 - Weight Resolution for Analysis; Display Mode; Analysis Number for Automatic Consecutive Counting; 4 Identification Codes (Ids)

Sartorius MA50
88 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
asap
(automatic searching/ automatic
programming): parameter determination
for semi-automatic mode
Automatic parameter determination
for shutoff criterion; semi-automatic;
percentual. In this case, you need to press
the Stop key to end an analysis; the
weight loss rate will then be calculated and
saved as a drying parameter.
Semi-automatic, absolute:
The analysis ends as soon as the weight
loss of a selectable time unit is less than
the defined limit, which can be selected
in milligrams. You will need to enter the
time unit and the weight loss.
Semi-automatic, in percent:
The analysis ends as soon as the weight
loss during a specified time period is less
than the defined limit, which can be select-
ed in percent of the initial weight. You will
need to enter the time unit and the weight
loss with reference to the initial weight.
Time:
The analysis ends as soon as the entered
time has elapsed.
Manual:
You must press the Stop key to end
an analysis.
33
Weight Resolution for Analysis
You can select the number of decimal
places to be displayed for the weight
resolution (MA100 only):
increased weight resolution during the
drying process only
output on the display and in the
printout
Display Mode
The following units can be selected for
displaying analysis results:
Moisture %L
decimal places, selectable
(with conversion factor
for MA100 only)
Dry weight %R
decimal places, selectable
Ratio %LR
decimal places, selectable
Weight loss mg
Residue g
Residue g/kg
Intermediate Results Printout
Intermediate results can be printed
either at user-definable time intervals or
by pressing the p key.
Analysis Number for Automatic
Consecutive Counting
The analysis number can be counted
automatically for all analyses performed
in sequence:
Press the Next soft key to save
this number
The number is set to 1 automatically
each time the analyzer is activated
Numbers are printed out in the header
for each analysis
Analysis with Former Sample
(100% function)
The sample weight of the last analysis is
calculated as 100% for the next analysis
4 Identification Codes (IDs)
Identification codes can be entered
for each drying program (such as Dairy
Farm, Berlin, powdered milk, Lot 1):
For user-defined printouts
Maximum of 20 characters for the
“Name” of 4 identification codes
(first part)
The related values (last part) can be
entered after pressing the ID soft key
during the analysis runs
Factory Setting
The drying programs can be reset to the
factory settings.
Searching for Programs
In the program memory
(after pressing the Prog. soft key):
Soft key q and Q
Enter the desired program number and
press the Number soft key
Enter the first or all characters of the
desired program number and press the
Name soft key
Statistics included (MA100 only)
Before activating the function “TARE:
Tare the sample pan”:
Enter the desired program number and
press the Prog. soft key
Weight
Moisture
%/mg
Time
Automatic Shutoff
/..s

Table of Contents

Related product manuals