Step Key (or instruction) Display/Printout
Otherwise, adjust the weighing system
Start soft key
This is displayed after adjustment:
--------------------
07.09.1999 13:04
Mod. MA100C
Ser. no. 90805355
Ver. no. 01-38-07
ID
--------------------
External calibration
W-ID
Nom. + 50.000 g
Diff. + 0.001 g
External adjustment
completed
Diff. 0.000 g
--------------------
07.09.1999 13:04
Name:
--------------------
7. Unload the weighing system
53
External Calibration/Adjustment
with a User-defined Weight
You can define a weight for calibration/
adjustment. External calibration/
adjustment must be performed with
weights that are traceable to a national
standard and that have error limits which
are at least 1/3 of the required tolerance
of the display accuracy. The defined
weight must equal at least 10% of the
maximum weighing capacity.
See page 52 for the external
calibration/adjustment sequence.
For this example, select “Ext. cal./adj.;
user-defined weight.”
The moisture analyzer has a
factory-set calibration weight value
(see “Specifications”).
Enter the factory-defined value
manually:
Select Setup menu: Press the s key
Select menu item
Device parameters
Select menu item Calibration/
adjustment
Change Cal./Adj. wt.
Tareo Start
WGH.SYS: Ext. cal. factory-def. wt
+ 50000g