Printout of the Analysis Parameters
– When the desired program is displayed:
Press the p key
> Printout (see example on the right)
Texts with more than 20 characters are
truncated.
Extra Functions
The following functions are available
in the program memory:
– Display programs
– Change the settings
– Copy or rewrite the programs by press-
ing the Copy soft key
– Load the programs by pressing the
Load soft key
– Print out the list with program numbers
and names
--------------------
13.07.1999 13:06
Mod MA100C
Ser. no. 90706913
Ver. no. 01-38-07
ID
--------------------
SETUP
ANALYSIS
--------------------
Prg. 1
--------------------
Program name
Program name:
RYE
Heating program
Standard drying
Temperature:
105 ’C
Standby temperature
Off
Bar graph for weigh
Inactivated
Start of analysis
With stability + a
Time delay:
2 sec
End of analysis
Fully automatic
Weight resolution f
1mg
Display mode
Moisture (%
Number of decimal
2 decimal places
Conversion factor
Factor:
1.0000
Print intermediate
Off
Analysis ID # with
No
Analysis w/formers
No
Identification #
ID1:
COMPANY:
ID2:
CITY:
ID3:
LOT:
ID4:
NAME:
--------------------
Function Keys (Soft Keys)
Info Intermediate results for
phase drying
Prog. Parameter setting for drying
programs
Stat. Statistics on the analyses of
the selected drying program
ID Data input for 4 identification
codes (IDs)
Mode Parameter settings: “Decimal
places,” “Weighing result,”
“Display mode,” “Intermediate
results printout”
Tare Tare sample pan
Load Load the selected drying pro-
gram as the current program
Copy Store the current parameter
settings in the selected
program location
34