SARTORIUS Master
pro
15
External Span Adjustment with Stored Weight
Activate the service mode (see page 11).
Press q and then press v to select the “user ext.adj.” menu item.
Tare the balance, if necessary, and then press q to start external span
adjustment.
Center the prompted calibration weight on the weighing pan.
Note: Depending on the menu code setting (under 1 10 X), either calibration/ad-
justment is made automatically (code 1 10 1: calibrate, then auto-adjust), or
the deviation from the last calibration is displayed, but no adjustment is
made until you press a key (code 1 10 2: calibrate, then manual adjust).
If the error exceeds the tolerance limits, press q to make the correspond-
ing adjustment, or press c to cancel the adjustment procedure.
The balance then returns to the SELECT mode.
Overwriting the Internal Calibration Weight
Note: As a rule, this procedure is only necessary after the balance has been opened
for service/repair work.
Activate the service mode (see page 11).
Perform external span adjustment (make sure you use a precise weight).
Press v (several times) to select the “det. int. wt.” menu item.
Tare the balance, if necessary, and press q to overwrite the calibration
weight.
The built-in calibration weight is applied automatically, measured and writ-
ten to the EEPROM in the balance processor.
Press c if you need to cancel this routine.
The balance then returns to the SELECT mode.
Calibration with a User-defined Weight
You can define any weight as the calibration weight.
Activate the INPUT mode to enter the weight value and then store it under the
cal./adj. wt.” menu item.
Activate the service mode (see page 11).
Perform external or internal span adjustment.
Press v to select the “det. ext. wt.” menu item.
Tare the balance if necessary, and then press q to start the calibration
routine.
The last calibration weight that was entered in the INPUT mode is dis-
played.
Center the calibration weight you wish to define on the weighing pan.
The weight value is displayed and stored.
The balance returns to the SELECT mode.
0000.00 g
S user EXT.adj selec.
Max5200 g d= 0.01 g
- 1000.00 g
S user EXT.adj
Max5200 g d= 0.01 g
0000.00 g
S user EXT.adj selec.
Max5200 g d= 0.01 g
0000.00 g
S BEt. int. wt. selec.
Max5200 g d= 0.01 g
0000.00 g
S BEt. .EXt. wt. selec.
- 1000.00 g
S BEt. .EXt. wt.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Master Pro LP5200P and is the answer not in the manual?

Sartorius Master Pro LP5200P Specifications

General IconGeneral
BrandSartorius
ModelMaster Pro LP5200P
CategoryScales
LanguageEnglish

Related product manuals