External Linearization
There are two ways to perform external linearization:
1. with the Sartorius service software (PSION Server or SARTOCAS Service
Software for PCs) - if the adjustment error is greater than ± 2%, or
2. as described in this manual, with the balance in the service mode.
External Linearization with Manual Input of Weight Values
If adjustment weights of sufficient accuracy are not available (at least one digit
greater than the display accuracy) or if no calibration certificate is available,
you can program the balance with the weight value including all necessary
decimal places.
If this is the case, follow the steps listed below:
Manual Input of the Linearization Weights
Press ] to turn on the balance.
The balance performs a self-test.
0.00 g is displayed (depending on balance model)
Activate the SERVICE mode:
Enter the service password: »202122«
Press ö.
Press
F1F1
F1F1
F1 to activate the INPUT mode.
Press
F2 F2
F2 F2
F2 to activate the “lin. weight 1” menu item and use the numeric
keys to enter the calibration weight to input the adjustment weight value,
including all necessary decimal points, for the first linearization weight in
accordance with the calibration certificate.
Press
F1 F1
F1 F1
F1 to store the weight value entered, and then press
F2F2
F2F2
F2 to proceed
to the “lin. weight 2” menu item.
Use the numeric keys to enter the calibration weight for the second linear-
ization weight, press
F1F1
F1F1
F1 to store, and then press
F2F2
F2F2
F2 to proceed to the
lin. weight 3” menu item.
Use the numeric keys to enter the calibration weight for the third lineariza-
tion weight, press
F1 F1
F1 F1
F1 to store, and then press
F2F2
F2F2
F2 to proceed to the “lin.
weight 4” menu item.
Use the numeric keys to enter the calibration weight for the fourth lineariza-
tion weight and press
F1F1
F1F1
F1 to store.
After defining all 4 linearization weights, press ö to exit the INPUT
mode.
The balance returns to the weighing mode.
La-021e.BMP
La-022e.BMP
La-023e.BMP
La-001e.BMP
La-016e.BMP
La-019e.BMP
La-020e.BMP
F2 F1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Master Pro LP5200P and is the answer not in the manual?

Sartorius Master Pro LP5200P Specifications

General IconGeneral
BrandSartorius
ModelMaster Pro LP5200P
CategoryScales
LanguageEnglish

Related product manuals