135
ID code characters I 
1
)
Stat Status
ID Identification (identifier)
L ID Weighing series no. (lot)
W ID Weight set number
Nom Exact calibration weight
(value)
S ID Sample ID
NUM Numeric input
T1
Application tare memory 1
N Net weight (T1 = 0)
N1 Net weight (T1  0)
Qnt Piece count
Prc Percent
nRef Reference sample quantity
pRef Reference percentage
wRef Reference piece weight
Wxx% Reference percentage weight
mDef Target value for animal 
weighing
Mul Calculated result 
in animal weighing
x-Net Result in animal weighing
x-Res Calculated result in animal
weighing
Res Result using equation 
(calculation)
Setp Target value for 
checkweighing
Min Lower limit for checkweighing
Max Upper limit for checkweighing
Time Time that a value was stored
Compxx Component no. xx in formula-
tion
Tot.cp Total weight in formulation
n Transaction counter
Total Sum of all values
Avg. Average in statistics
s Standard deviation
srel Variation coefficient
Diff Difference between maximum
and minimum
Example: output weight of + 111.25507 mg
Position 1 2345678910111213141516
+111. 25507*mg*CRLF
Position 1: Plus + or minus sign – or space
Position 2: Space or weight value digit
Positions 3–10: Weight with decimal point; leading zeros are output as spaces
Position 11: Space
Positions 12–14: Characters for unit of measure or space
Position 15: Carriage return
Position 16: Line feed
Data Output Format with 22 Characters
When data with an ID code is output, the ID code consisting of 6 characters precedes the
data with the 16-character format.  These 6 characters identify the subsequent value.
12345678910111213141516171819202122
IIIIII+DDDDDDDDD * UUUCRLF
*****–......... ***
**********
I: ID code character
1
) U: Unit symbol
1
) see “Toggle
*: Space between Weight Units”
D: Digit or letter CR: Carriage return
LF: Line feed
1
) depends on balance type; e.g., not all units and characters are 
available on balances verified for use in legal metrology
Special Codes
12345678910111213141516171819202122
Stat****** * *––***** *CRLF
HH
LL
C
*: Space L: Underload
– –: Final readout mode 
(unstable weight) L L: Underload in checkweighing
H: Overload C: Calibration/adjustment
H H: Overload in checkweighing Draft shield and ionizer status; similar
to data output format with 16 characters
Error Codes
12345678910111213141516171819202122
Stat*****Er r *###****CRLF
*: Space # # #: Error code number