13
§ Plug power cord into the AC adapter
§ To power the balance with AC current, plug the power cord into
a wall outlet (mains supply)
Charging the Rechargeable Battery for Saving Data:
All data is saved in the battery-backed memory. When initially operating the balance,
leave it connected to AC power for one day to charge the battery. When the balance
is disconnected from AC power, the balance-generated data will remain stored for
approximately three months. In the standby mode, data is retained in the memory via
the power supply. Be sure to print out data before storing your balance for a relatively
long period.
Safety Precautions
The AC adapter rated to Class 2 can be plugged into any wall outlet
without requiring any additional safety precautions. The ground or earth terminal is
connected to the scale housing, which can be additionally grounded, if required.
The data interface is also electrically connected to the balance housing (ground).
Note:
This equipment has been tested and found to comply with the limits for a Class A digital
device, pursuant to Part 15 of the FCC rules. These limits are designed to provide
reasonable protection against harmful interference when the equipment is operated in
a commercial environment. This equipment generates, uses and can radiate radio
frequency energy and, if not installed and used in accordance with the instruction
manual, may cause harmful interference to radio communications. Operation of this
equipment in a residential area is likely to cause harmful interference in which case the
user will be required to correct the interference at his own expense. Changes or
modifications not expressly approved by Sartorius AG could void the user‘s authority
to operate the equipment.
Connecting Electronic Peripheral Devices
§ Make absolutely sure to unplug the balance from AC power before you connect or
disconnect a peripheral device (printer or PC) to or from an interface port
Warmup Time
Each time you move your balance to another location, you must condition it for at least
12 hours to the new location. To deliver exact results, the balance must warm up for at
least 2 hours after initial connection to AC power. Only after this time will the balance
have reached the required operating temperature.
Using Balances Verified as Legal Measuring Instruments in the EU*:
$ The balance must warm up for at least 24 hours after initial connection
to AC power
$ Warmup time each time power is turned on the ME614S-0CE, ME415S-0CE,
ME414S-0CE: at least 30 minutes
$ Always wait for the power-on adjustment routine to be completed
(for requirements see page 55).
$ isoCAL function switched off: see information on page 55.
* including the Signatories of the Agreement on the European Economic Area

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius ME5-F and is the answer not in the manual?

Sartorius ME5-F Specifications

General IconGeneral
BrandSartorius
ModelME5-F
CategoryScales
LanguageEnglish

Related product manuals