143
Error Code/Message Displayed Cause Solution/Remarks
Differential weighing:
SAMPLE: Confirm delete/ “SAMPLE: delete/omit” prompt Select Yes to delete
omit on display page for samples Select Omit to omit
SAMPLE: Include “SAMPLE: include” prompt Select Omit to include
on display page for samples a sample already omitted
Cannot store File manager: Delete lot(s)
Not possible to save data
No available memory
Cannot load File manager: Delete lot(s)
Not possible to load data
Memory capacity limit reached
Only 30 backweighs An attempt was made to save a None
possible 31st backweighing operation
LOT: already exists Lot already exists on the display Choose a different lot ID
page for LOTS
No sample
The Sample key was pressed when Save sample first
the display page for lots was shown,
but there are no samples in the lot selected
Out of range On the display page for LOTS Enter the correct lot or sample ID
or
SAMPLES, an alphanumeric
lot or samle ID was input and not found
Not enough memory space An attempt was made to save more Use less memory or delete one or more lots
or than 999 samples using the
999 samples maximum numeric keys and the #Spl key
Sample omitted An attempt was made to save None
data from an omitted sample
Value too small to accept An attempt was made to Place the particular weight on the balance
save a tare, initial or backweight
that is less than one display digit
No choice available Factor was selected while attempting Selection not possible
to activate the display page for results
No 2
nd
resolution available Contact your local Sartorius Service Center
CF not possible Only one sample or certain portions Samples can be deleted one
of a sample can be deleted by pressing at a time on the display page for samples
the c key. This message indicates that
further delete functions are not possible.
Calculated statistics Message output when statistics are being Goes out automatically
calculated. This process can take several
seconds if there are many samples.
No statistics available No valid backweights available in this lot Goes out automatically
No net initial wts. In serial and combined weighing, Measure initial weights
available no initial weights found
If any other errors occur, contact your local Sartorius Service Center!