73
Example
Buoyancy: Determine the Density of Samples of a Solid Using the Buoyancy Method. Reference Liquid: Water
Settings (changes in the factory settings required for this example):
Setup: Application parameters: Application 1: Density
Step Press key(s) (or follow instructions) Display/Output
1. Delete previously stored values c
if necessary
2. Change parameters,
Param. soft key
if necessary
3. Position the sample holder
(immerse in water)
4. Tare the balance t
5. Determine the weight of the
sample in air: place sample
on the weighing pan
6. Store weight value
Wa soft key
7. Determine the weight of the
sample in liquid: place sample
in the sample holder
8. Store weight value in liquid
Wfl soft key
Density of sample is displayed
oo Q ql
DENSITY:
Temperature : Temp + 20.0 C
Buoyancy corr: Corr + 1.000000
Air density : LA + 0.001200 g/
g
Max 210 g
0%
DENSITY: Temp=20.0 C
C
L
S
T
d=0.01mg
100%
000000
F
S
a
s
Cal WaParam.
E
H
N
g
Max 210 g
0%
DENSITY: Temp=20.0 C
C
L
S
T
d=0.01mg
100%
000000
F
S
a
s
Cal WaParam.
E
H
N
g
Max 210 g
0%
DENSITY: Temp=20.0 C
C
L
S
T
d=0.01mg
100%
1502041
F
S
a
s
Cal WaParam.
E
H
N
g
Max 210 g
0%
DENSITY: Temp=20.0 C
C
L
S
T
d=0.01mg
100%
1502041
F
S
a
s
Cal WflParam.
E
H
N
g
Max 210 g
0%
DENSITY: Temp=20.0 C
C
L
S
T
d=0.01mg
100%
1324084
F
S
a
s
Cal WflParam.
E
H
N
g
Max 210 g
0%
DENSITY: calculated density
C
L
S
T
d=0.01mg
100%
838
F
S
a
s
Cal StartParam. Vol.
E
H
N
g/
cm
3

Table of Contents