Operating Instructions Microsart
®
Manifold 19
Installation
4.2 Selecting an Installation Site
t Make sure that the following conditions are met at the installation site:
Condition Features
Ambient conditions Suitability tested (see Chapter “13.5 Ambient
Conditions at the Installation Site”, page 40 for
ambient conditions)
Setup surface Stable and even surface, as well as sufficient
operating height
Sufficient surface area for the product and the
peripheral devices (for the product space
requirements, see Chapter “13.1 Dimensions and
Weights”, page 38; for peripheral device space
requirements, see instructions for the peripheral
devices, see Chapter “1.2 Accompanying
Documents”, page 7)
Access to a vacuum source and a drain is provided
Access to parts relevant to
operation
Convenient and safe
Manual_Microsart_Manifold_W166001-e.indb 19 28.02.19 10:03