Operating Instructions Microsart
®
Manifold 29
Cleaning and Maintenance
Interval Component Action Chapter,
page
Weekly /
depending on
application
Manifold Autoclave the manifold. 7.6, 32
Annually O-ring The O-rings shift. 7.4, 30
Only in the event
of leakage /
damage
Silicone foot Replace the silicone feet. 7.7, 32
7.3 Cleaning the Manifold
7.3.1 Cleaning after each Filtration:
Procedure
t Spray the base support with disinfectant or fill up to a few milliliters.
t Vacuum out the disinfectant.
t Rinse the manifold with sterile water.
7.3.2 Cleaning at the End of the Work Day
Procedure
t Spray the base support with disinfectant or fill up to a few milliliters.
t Allow the disinfectant to take effect for 3 5 minutes.
t Vacuum out the disinfectant.
t Rinse the manifold with sterile water.
t Empty the manifold by tipping it (see Chapter “9 Decommissioning”, page 35).
t Dry the sintered frits using lab cloths.
Manual_Microsart_Manifold_W166001-e.indb 29 28.02.19 10:03