84 Installation and Operating Instructions Secura, Quintix, Practum
Technical Specifications
Practum
®
Model 224-1x
2
) 124-1x
2
) 64-1x
2
) 513-1x
2
) 313-1x
2
) 213-1x
2
) 3102-1x
2
) 2102-1x
2
)
Accuracy class
kkkKKKKK
Type
3
) SQP-A SQP-A SQP-A SQP-B SQP-B SQP-B SQP-C SQP-C
Max g 220 120 60 510 310 210 3,100 2,100
Scale interval d mg 0.1 0.1 0.1 1 1 1 10 10
Verification scale interval e mg 111101010100100
Min mg 10 10 10 20 20 20 500 500
Min (only for Models ...-1OIN) mg 100 100 100 200 200 200 5,000 5,000
Tare (subtractive) ) 100% of the max. weighing capacity
Typical stabilization time s 1 2 2 1 1 1 1.5 1.5
Weighing pan size mm d 90 d 90 d 90 d 120 d 120 d 120 d 180 d 180
Weighing chamber height* mm 209 209 209 209 209 209
Net weight, approx. kg 4.5 4.5 4.5 4.9 4.9 4.9 3.1 3.1
Practum
®
Model 1102-1x
2
) 612-1x
2
) 6101-1x
2
) 5101-1x
2
)
6100-1x
2
) 5100-1x
2
)
Accuracy class
KKKKKK
Type
3
) SQP-C SQP-C SQP-E SQP-E SQP-E SQP-E
Max g 1,100 610 6,100 5,100 6,100 5,100
Scale interval d mg 10 10 100 100 1,000 1,000
Verification scale interval e mg 100 100 1,000 1,000 1,000 1,000
Min mg 500 500 5,000 5,000 50,000 50,000
Min (only for Models ...-1OIN) g 5 5 50 50 50 50
Tare (subtractive) ) 100% of the max. weighing capacity
Typical stabilization time s 1.5 1.5 1.5 1.5 1 1
Weighing pan size mm d 180 d 180 d 180 d 180 d 180 d 180
Net weight, approx. kg 3.1 3.1 3.1 3.1 3.1 3.1
2
) Possible terms for country-specific models:
x = CEU: Verified balances with EC Type Approval Certificate D12-09-014 without country specific additions
x = CFR: Verified balances with EC Type Approval Certificate D12-09-014 for France only
x = CIT: Verified balances with EC Type Approval Certificate D12-09-014 for Italy only
x = CCH: Verified balances with EC Type Approval Certificate D12-09-014 for Switzerland only
x = CN: CMC Type Approval Certificate for China
x = OJP: Balance with Type Approval Certificate for Japan
x = OBR: Balance with Type Approval Certificate for Brazil
x = ORU: Balance with Type Approval Certificate for Russia
x = OIN: Balance with Type Approval Certificate for India
x = OAU: Balance with Type Approval Certificate for Australia
3
) All models with “...CN”: type “SQP”
* Upper edge of the weighing pan to the lower edge of the upper draft shield panel
Model-specific Data