80 Installation and Operating Instructions Secura, Quintix, Practum
Technical Specifications
Secura
®
Model 224-
1x
1
)
124-
1x
1
)
1103-
1x
1
)
613-
1x
1
)
513-
1x
1
)
313-
1x
1
)
213-
1x
1
)
6102-
1x
1
)
5102-
1x
1
)
3102-
1x
1
)
2102-
1x
1
)
1102-
1x
1
)
612-
1x
1
)
6101-
1x
1
)
3101-
1x
1
)
Weighing capacity g 220 120 1,100 610 510 310 210 6,100 5,100 3,100 2,100 1,100 610 6,100 3,100
Readability mg 0.1 0.1 1 1 1 1 1 10 10 10 10 10 10 100 100
Repeatability
(standard deviation)
mg 0.10.1111111010101010105050
Linearity mg 0.2 0.2 2 2 222202020202020100100
Min. initial weight
according to USP, typical
g 0.12 0.12 1.5 1.5 1.5 1.5 1.5 12 12 12 12 12 12 60 60
Sensitivity drift
between +10 to +30°C
± ppm/K 1.5 1.5 1.5 2 22222222222
Typical stabilization time s 2 2 1.5 1 11111111111
isoCAL Settings:
– temperature change
– time interval
K
h
1.5
4
1.5
4
1.5
4
2
6
2
6
2
6
2
6
2
6
2
6
2
6
2
6
2
6
2
6
2
6
2
6
Display result (depending
on the filter level)
s 0.2 0.2 0.1/
0.2
0.1/
0.2
0.1/
0.2
0.1/
0.2
0.1/
0.2
0.1/
0.2
0.1/
0.2
0.1/
0.2
0.1/
0.2
0.1/
0.2
0.1/
0.2
0.1/
0.2
0.1/
0.2
Weighing pan size mm d 90 d 90 d 120 d 120 d 120 d 120 d 120 d 180 d 180 d 180 d 180 d 180 d 180 d 180 d 180
Weighing chamber height* mm 209 209 209 209 209 209 209 –––––
Net weight, approx. kg 5.1 5.1 5.9 5.1 5.1 5.1 5.1 5.2 5.2 5.2 5.2 5.2 5.2 5.2 5.2
Quintix
®
Model 224-
1x
1
)
124-
1x
1
)
64-
1x
1
)
613-
1x
1
)
513-
1x
1
)
313-
1x
1
)
213-
1x
1
)
6102-
1x
1
)
5102-
1x
1
)
3102-
1x
1
)
2102-
1x
1
)
1102-
1x
1
)
Weighing capacity g 220 120 60 610 510 310 210 6,100 5,100 3,100 2,100 1,100
Readability mg 0.1 0.1 0.1 11111010101010
Repeatability
(standard deviation)
mg0.10.10.111111010101010
Linearity mg 0.2 0.2 0.2 22222020203030
Sensitivity drift
between +10 to +30°C
± ppm/K 1.5 1.5 1.5 333333344
Typical stabilization time s 22211111111.51.5
isoCAL Settings:
– temperature change
– time interval
K
h
1.5
4
1.5
4
1.5
4
2
6
2
6
2
6
2
6
2
6
2
6
2
6
2
6
2
6
Display result (depending
on the filter level)
s 0.2 0.2 0.2 0.1/0.2 0.1/0.2 0.1/0.2 0.1/0.2 0.1/0.2 0.1/0.2 0.1/0.2 0.1/0.2 0.1/0.2
Weighing pan size mm d 90 d 90 d 90 d 120 d 120 d 120 d 120 d 180 d 180 d 180 d 180 d 180
Weighing chamber height* mm 209 209 209 209 209 209 209 –––––
Net weight, approx. kg 4.9 4.9 4.9 4.9 4.9 4.9 4.9 5.2 5.2 5.2 4.7 4.7
Model-specific Data