Example:
Weighing in 3 components of a formulation recipe.
Settings (changes in the factory settings required for this example):
Application parameters: Application: Net-total formulation
Setup: Printout:
PrtProt: 7.6. “Component printout: Print when value stored;”
then select the menu line item of your choice (see “Configuration” for options)
Setup: Printout:
PrtProt: 7.6.: “Total printout: Print when CF pressed;”
then select the menu line item of your choice
1 Place empty container on the platform
)
2 Tare the scale
Note: If the automatic tare function is enabled, you do not need to press the ) key to tare the scale;
the tare weight is saved automatically when you place the container on the platform
Prompt to fill and save the first component is shown
3 Add the first component to the container (in this example, 1100 g)
The weight of the first component is displayed
O
4 Store the weight of the first component
Cmp001+ 1.100 kg The component weight is printed automatically
Application 3: Net-total Formulation R