36
Anz-00F.eps
Anz-500F.eps
Span Adjustment
Note: There are two methods for adjusting the span:
1. Using the service software
2. As described in the following:
– Select menu item 1  5  1, „Calibration/adjustment:  Accessible“ in the
operating menu (factory setting). Make sure the menu access switch is open
(see page 19).
– If necessary, select the calibration weight unit in the operating menu
(item 1  4  x; factory setting: 1  4  2*, „kg“).
– Unload the balance/scale (in this example, model TE6101) and press t to
zero it, if necessary.
– Press and hold the t key (> 3 sec) to initiate the calibration routine.
– The required calibration weight is displayed without a weight unit; e.g.,
„+ 5000.0“
– Center the displayed weight* on the weighing pan/load plate
(in this example, 5000.0 g).
Important: The balance/scale will only accept a weight that is within a tolerance range
of approximately 2% of the nominal weight.  Greater deviations can be
corrected only using the service software.
– After calibration, the readout shows the calibration weight* with weight unit
(in this example, „+ 5000.0 g“).
– Remove the calibration weight.
– Check the span and perform adjustment again if necessary.
      * depends on model
Anz-5000gF.eps
Anz-151F.eps