6
General Notes
Service Guide
1. Check the repeatability
Balance/Scale with strain gauge system (standard deviation, see page 34)
Balance/Scale with double lever system (standard deviation, see page 39)
2. Check the off-center loading error and adjust as required
Balance/Scale with strain gauge system (see page 35)
Balance/Scale with double lever system (see page 40/41)
3. Adjust the sensitivity with external calibration weight
Balance/Scale with strain gauge system (see page 36)
Balance/Scale with double lever system (see page 46)
4. Adjust the linearity is preferably performed using the sequential method,
or with calibrated weights,
Balance/Scale with strain gauge system (see page 37)
Balance/Scale with double lever system (see page 47)
Activating the BPI Mode/ Perform only before working with the Sartorius Service Software;
Deactivating write-protect for calibration/adjustment. Linearization and programming a data record when
exchanging the main PCB.