Getting Started
Connecting the Equipment to AC Power
Danger!
Before connecting the equipment to power, make sure the power cords are not live.
The main switch must be in the “OFF” or “0” position.
Installation
For the electrical connection to the power supply, a feeder with a cross section
of 3 + 1.5 mm
2
is required.
The supply voltage is 230 V (±10%) at 50 Hz (±5%). Before connecting the feeder, read
and follow all safety instructions in these operating instructions and observe the data on
the manufacturer’s ID label. To prevent interference from other devices (e.g., from feed-
back), we recommend installing a separate line and securing it with a 6A semi-time lag
circuit breaker. We also recommend installing additional safety devices within the electri-
cal circuit, such as an emergency shutoff switch.
Caution:
When you install data cables, make sure that power cables and data cable are not parallel,
as this can lead to inductive interference.
§ Use the key supplied to open the control cabinet.
§ Feed the cable into the cabinet from below.