Motor Cable
The motor cable is attached to the checkweigher frame at the factory.
This must not be modified.
§ Check whether the motor cable is unobstructed and not in contact with any objects (e.g.,
the load plate on the checkweigher).
§ Check whether the plug on the motor cable is firmly connected and the closing clamp for
the connection is in the “up” position.
Switching on Power/Initial Commissioning
Before initial commissioning:
§ Check whether the local supply voltage matches the rating listed on the manufacturer’s
ID label.
§ Connect the equipment to power.
§ Check the supply voltage, phase, neutral connector and ground connection. Check the
protective earth (grounding) in accordance with national regulations.
§ Remove all objects from the conveyor belt before turning on the equipment.
§ Turn the main switch to the “1” or “ON” position. The checkweigher performs a self-test.
Wait until “0” is displayed on the weight readout.
§ Turn the “ON/OFF” switch to the “ON” position. The conveyor belt turns.