Example:
“Calibration/Adjustment” Function via RS-232 Interface
Purpose
Calibration is the determination of any difference between the measured value displayed
and the true weight (mass) of a sample. Adjustment is the correction of this difference,
or its reduction to a suitable level within maximum permissible error limits.
Characteristics
The adjustment procedure should only be started when
The weigh cell is not loaded
The weigh cell is tared
The weighing signal is stable
The sensitivity of the balance can be corrected by max. 2%.
If these criteria are not met, error message “Err 02” appears.
Error message “Err02”:
Note ambient conditions
Weigh cell needs stability
If necessary, change the pre-configured balance parameters:
Select Ambient conditions menu item 1.1.1.4 (very unstable) or execute interface
command ESC N
Adjustment can be made using different weight units:
CAL.Unit > Gram, Kilogr.