Power On Switch
• The ENABLE/LOCK button is the main On/Off
power switch. When depressed it will first lock
the table to the floor, then display the table posi-
tion, Normal or Reverse. You have approximately
10 seconds to activate a function or positioning
button before the hand control automatically turns
off.
Disable
• The
DISABLE
button, when depressed, turns off
the positioning buttons on the hand control and
will stop the movement of a function when the
foot control is being used. To re-activate the but-
tons on the hand control you must first depress
the
ENABLE/LOCK
button and then the desired function button
Table Position
• You must choose how you want
the table position to be config-
ured by pressing either
Normal
,
(head over column) or
Reverse
,
(head opposite end of column)
Position
. The position activated
will affect certain functions:
1. Flex or Reflex cannot
be activated when in Reverse Position.
2. Trendelenburg, Reverse Trendelenburg, Tilt Left and Tilt Right are automati-
cally configured to operate in conjunction with the Normal and Reverse Posi-
tions. Refer to the Controls & Indicators section.
3. Reverse position is not
available on the Foot or Emergency Override Con-
trols even though the hand control has been configured to operate in the
Reverse Position.
WARNING
It is strongly recommended that the
DISABLE
button be de-
pressed to switch the hand controller OFF
after table posi-
tioning has been completed. This will prevent accidental activation of a
function during procedures.